PRODUCT DETAILS
The AR2 Urine Test Cup from Test Cup Innovations has been designed and manufactured to offer compliance with the Australia and New Zealand Drug Standard, ASNZ 4308:2008. High quality components and leading manufacturing techniques has reduced the potential for false positives and false negatives. The AR2 Urine Test Cup provides preliminary testing for eight drugs of abuse.
- Amphetamines
- Benzodiazepines
- Cocaine
- Synthetic Marijuana
- Methamphetamines
- Opiates
- Oxycodone
- THC (Cannabis)
As a two piece test cup the donor does not come into contact with the testing panel component of the AR2 Urine Test Cup device, significantly reducing the potential risk for tampering. The AR2 also features in-built temperature and adulteration panels to provide an indication of sample tampering.
PROCEDURE
2. Record temperature and verify it is within range (32 - 38°C )
3. Remove lid from packaging; align the lid on the cup and screw closed. Ensure correct alignment to prevent cross-threading and leakage.
4. Place cup on its side. Negative results of drug screen test can be interpreted as soon as the control lines appear and there are visible Test lines which usually occur within 1 minute. Positive results of drug screen test should be read at 5 minutes. All results remain stable for 60 minutes.